User's Manual

Revision B ApexPro Telemetry System 2-13
2001989-134
Safety: For Your Safety
Classification
The telemetry system is classified, according to IEC 60601-1, as:
Underwriters Laboratories, Inc.
Medical Equipment
With respect to electric shock, fire and mechanical hazards
only in accordance with UL 2601-1, and CAN/CSA C22.2
NO. 601.1 and if applicable, IEC 60601-2-27, IEC 60601-2-
30, and IEC 60601-2-49.
Type of protection against electrical shock Transmitter — Internally powered
Receiver system — Class I
Degree of protection against electrical
shock
ApexPro Transmitter — Type B applied part
ApexPro CH Transmitter—Type CF Defibrillation proof applied part (not for sale
outside of the U.S.)
Degree of protection against harmful
ingress of water
ApexPro Transmitter — IPX3 (IEC 60529)
ApexPro CH Transmitter — IPX7 (IEC 60529)-not for sale outside of the U.S.
Receiver system — Ordinary Equipment (enclosed equipment without protection
against ingress of water)
Degree of safety of application in the
presence of a flammable anesthetic
mixture with air or with oxygen or nitrous
oxide
Equipment not suitable for use in the presence of a flammable anesthetic mixture
with air or with oxygen or nitrous oxide
Method(s) of sterilization or disinfection
recommended by the manufacturer
Not applicable
Mode of operation Continuous operation