User Manual

Table Of Contents
Revision A - Draft 7 APEX, ApexPro, and ApexPro CH Transmitter Programming Instructions 25
2001989-135
Apex Transmitter
To set the band:
1. Select the Set Band button, reference C in “Program Apex Transmitter
Window” on page 23.
2. Another window appears. Select the correct band. (Selecting one
eliminates the others.) Select OK.
3. The change is shown on the Program Apex Transmitter window.
Reference Lead Setting
For 5- or 6-Lead Cables
The reference lead setting has no effect when 5- or 6-lead cables are used.
For 5- or 6-lead cables, the reference lead defaults to RL.
For 3-Lead Cables
For 3-lead cables, one of the lead wires (LL, LA, or RA) is used to connect the
reference voltage to the patient. Selection of the reference lead determines
which ECG waveform will be displayed at the CIC, according to the table
below.
805A
Select this
reference lead...
to view this
waveform at the CIC
LL (left leg) Lead I
LA (left arm) Lead II
RA (right arm) Lead III
Draft