User Manual

Table Of Contents
Revision A - Draft 7 APEX, ApexPro, and ApexPro CH Transmitter Programming Instructions 45
2001989-135
ApexPro CH Transmitter
return to the Select Download Type window, or Cancel to exit the update
process.
To download the Service code, proceed with step 5 in “Erase &
Download Service Code” on page 45
or
Select Cancel to exit the update process and return to the Program
ApexPro CH Transmitter window. (Selecting Cancel closes Super
User functions.
10. Confirm that all Power-up Self Test Status messages indicate Passed. See
“Appendix 1: ApexPro CH Troubleshooting” on page 49 if any failures
occur.
Erase & Download Service Code
1. Access the high-level technical functions of the software as described in
“Access Super User Mode” on page 52.
2. The Program ApexPro CH Transmitter window will display additional
features, including a Download EEPROM Code button (reference E in
“Appendix 2: Access ApexPro CH Technical Functions (Super User)” on
page 52.)
3. Select the Download EEPROM Code button. The message shown below
will display.
NOTE
NOTENOTE
NOTE
If the above message does not display, you must manually reset the
program as instructed in “Reset the Transmitter Manually” on page 51.
4. Using a small pointed device (such as the tip of a pen) press in and hold
the Clear Transmitter Memory button in the side of the Apex
programming device. Another window will display after approximately 4
seconds. Or select Cancel to stop the download process and return to the
Super User window.
391A
386A
Draft