Table Of Contents
Revision A - Draft 7 APEX, ApexPro, and ApexPro CH Transmitter Programming Instructions 49
2001989-135
ApexPro CH Transmitter
ApexPro CH Appendices
Appendix 1: ApexPro CH Troubleshooting
LED Status Problems
Condition Possible Cause Possible Actions
No lights are blinking. Hardware is not connected
properly.
Check the hardware connections as described in “Connect the Hardware” on
page 11.
Software is not installed. Install the Apex & ApexPro Tx Config software as described in “Install
the Programming Device” on page 9.
Software is not running. Start the Apex & ApexPro Tx Config software as described in “Run
the Apex & ApexPro Programming Box Software” on page 13.
Batteries do not have enough
power.
Replace the batteries:
1. Disconnect the transmitter from the programming device.
2. Replace the transmitter batteries with two new AA alkaline batteries.
3. Reconnect the transmitter to the programming device.
One or both firmware files are
corrupt.
Follow the instructions in “Update Transmitter Firmware” on page 42.
All lights blink rapidly several
times and then slowly two
times.
Normal status when hardware
has just been properly
connected.
No action needed.
All lights blink on and off every
second.
Normal status when hardware
is properly connected and the
software is running.
No action needed.
RA and LA LED lights are
blinking.
You have chosen to download
an Application or Service file.
Follow the instructions in “Update Firmware Code” on page 42.
Transmitter is receiving a
download of application
firmware.
No action needed.