ApexPro Transmitter
To change the setting:
1. Select the Set Lead button, reference C in “Program ApexPro
Transmitter Window” on page 59.
2. Another window appears. Select the correct lead. (Selecting one
eliminates the others.) Select OK.
3. The change is shown on the Program ApexPro Transmitter window.
Alarm Pause Setting
. The purpose of the alarm pause setting is to allow clinicians to adjust the
transmitter and/or the patient without setting off alarms. The alarm pause
setting is an amount of time, in minutes, during which alarms will not
activate. After this time has elapsed, the alarm pause will automatically be
deactivated and the transmitter will resume normal operation. You can
manually turn off the alarm pause at any time by pressing Verify Leads
and Graph at the same time.
NOTE
NOTENOTE
NOTE
Activation and deactivation of the alarm pause is done on the transmitter
itself. The setting in the software only sets the number of minutes to wait
until the alarm pause times out.
To change the setting:
1. Select the Set Pause button, reference D in “Program ApexPro
Transmitter Window” on page 59.
2. Another window appears.
3. Set the correct number of minutes. Either use the up- and down-arrows
to scroll to the desired number or highlight and type in the number to
use.
4. Select OK.
5. The change is shown on the Program ApexPro Transmitter window.