72 APEX, ApexPro, and ApexPro CH Transmitter Programming Instructions Revision A - Draft 7
2001989-135
ApexPro Transmitter
Verify Correct Operation
Verify the Transmitters Firmware Versions
1. Disconnect the transmitter from the programming device and wait for
the transmitter to reboot. (All the LED lights flash rapidly several times
and then slowly two times.)
2. Reconnect the transmitter to the programming device.
3. When the transmitter’s settings reappear in the Program ApexPro
Transmitter window, verify that the Code Version App: text string
matches the .app file name listed on the data disk. Also verify that the
Code Version Mfg: text string matches the .mfg file name listed.
4. Confirm that all Power-up Self Test Status messages indicate Passed.
Call Technical Support (page 78) if any failures occur.
Verify Transmitter Operation
Complete the checkout procedures identified in the ApexPro Telemetry
Transmitter Service Manual.
Track the Software Upgrade (Field Engineer Use Only)
For this upgrade, complete the following.
1. Use the MDOC number found on the data disk as the tracking number in
the PROACTIVE REPAIR# window of the Clarify case.
2. Bill travel and labor only as authorized by Marketing - CIC/ApexPro
using the MDOC number as the PO#.
3. Fill out and fax the “Update Installation Verification Form” provided
with this manual. (Please make additional copies as needed.)
460A