Operating Instruction

Simplified Operating Instruction, 340M OB Telemetry, 608-614MHz
GE Medical Systems Information Technologies 4/9/01
Page 3 of 8
Side:
(4) Gray Cables total
- Gray cable, no plug:
Not used
-Gray cable, 5700 Transducer at end, nothing beyond
Not used
-Gray cable, 5700 Transducer at end, mechanically face connected to slave 5700
transducer
This slave 5700 plug connects to Ultrasound input jack of the 342 transmitter.
-Gray cable, round 12 pin white connector at end
This plugs into the UA connector jack of the 342 Transmitter.
325 Maternal/Fetal Simulator (for use at ITS, Menlo Park, SAR):
Front Panel:
FECG/MECG section:
RATE bpm:
Set to 120
QRS AMPLITUDE:
Set to 2000
CMR:
Set to BAL
MODE:
Set to FECG
QRS POLARITY:
Set to (+)
ULTRASOUND/FMD section:
RATE bpm:
Set to 120
SIGNAL LEVEL:
Set to HIGH
MODE:
Set to US/FMD
UA Section
Black Slide Switch: