Step-by-step Instructions

4-10 Model 340 Telemetry System Revision A
2006899-001
Setup Procedures: Performing a Functional Checkout
Performing a Functional Checkout
Initial Conditions
Turn on the transmitter, the receiver, and the monitor attached to the
receiver.
Figure 4-13. Applying Power
Testing the Radio Frequency
1. Check the status of the Signal indicator on the receiver:
u
Continuous Green: indicates the transmitter is active and the
batteries have adequate capacity.
u
Flashing Green: indicates the signal strength is weak or
marginal.
2. Check the status of the
Battery indicator on the receiver:
u
Off: the transmitter batteries have power.
u
Flashing Red: the transmitter batteries are low and should be
replaced before further patient use.
u