Product Manual

1-12 SEER
®
Light 2040604-012D
Manual Information
Underwriters Laboratories, Inc.
Classified by Underwriters Laboratories Inc. with respect
to electric shock, fire, mechanical and other specified
hazards, only in accordance with UL 60601-1, CAN/CSA
C22.2 No. 601.1, EN/IEC 60601-1, and IEC 60601-2-47.
Information Technology Equipment
The hardware components supplied by GE for the MARS
®
Holter
analysis workstation, on which the SEER Light Connect application
runs, are considered to be Information Technology Equipment (ITE).
These individual components have been found to comply with the
standard for Safety of Information Technology Equipment, including
Electrical Business Equipment EN/IEC/UL 60950.
The software used in the MARS
®
Holter analysis workstation is
considered as medical software. The software has been designed and
manufactured to the appropriate medical regulations and controls.
In order for the MARS
®
Holter analysis workstation to comply with
medical equipment leakage current requirements, a medical grade
uninterruptible power supply (UPS) must be used (EN/IEC/UL 60601-1)
to power all non-medical equipment.
In addition, non-medical electrical equipment must comply with EN/IEC/
UL safety standards that are relevant to that equipment (i.e., IEC 60950,
Safety of Information Technology Equipment).
Equipment Symbols
The following symbols may appear on the equipment.
Attention. Consult accompanying documents.
Manufacturer name and address.