Product Manual

2-16 SEER
®
Light 2040604-012D
Equipment Overview
Battery Disposal
Do not burn the batteries, or dispose of batteries with other medical
wastes. Contact your local recycling center for information on proper
disposal of used batteries.
Care and Cleaning
Follow these cleaning instructions for the recorder, controller and the
connect device.
Keep the device clean between each use to prevent infection.
Remove the batteries and disconnect any power and USB cords
before cleaning the device.
Use a piece of cloth dampened with alcohol to clean the device and
the patient cable.
NOTE
Do not clean and reuse electrodes. Use new electrodes for each
use.
The devices cannot be sterilized.
Do not immerse the device in any liquid, or allow liquids to enter
through connections, ports or buttons.
Do not use xylene or petroleum-based solvents for cleaning the
device.
In case of a malfunction, call a service technician and precisely
describe the problem.
Check the patient cable and connectors every month by connecting to
an ECG simulator.
Storage and Operating Conditions
Follow these guidelines when storing and operating the SEER Light
devices.