Product Manual
3-2 SEER
®
Light 2040604-012D
Preparing the Patient
Placing Electrodes
Two Channel, Five Electrode
In the two channel, five electrode configuration, two channels of ECG
data are bipolar. Red positive (+) is referenced to white negative (–) and
brown positive (+) is referenced to black negative (–).
045A
AHA
Color
IEC
Color
Channel Lead Location
A Red Yellow CH 1(+) II (+) Fifth intercostal space,
anterior axillary line.
B White Red CH 1 (-) II (-) Below the right clavicle, just
lateral to the midclavicular
line.
C Brown Green CH 2 (+) mV1 (+) Fourth intercostal space, right
sternal edge.
D Black White CH 2 (-) mV1 (-) Below the left clavicle, just
lateral to the midclavicular
line.
E Green Black Ground On the lower right chest wall,
on a rib.
B
D
A
E
C