Product Manual

4-2 SEER
®
Light 2040604-012D
Entering Patient Data
Entering Patient Data
Patient data can be entered via the controller or the connect device.
WARNING
MIXING PATIENT DATA — Accidental mixing of
patient data can result in misdiagnosis and incorrect
treatment.
Always transfer or delete old data before beginning a new
recording session, and promptly enter patient
demographics.
Using the SEER Light Extend Controller
Be sure that all information from previous patient recordings has been
transmitted to the controller.
If data remains on the recorder that has not been transmitted to the
controller, the controller will display the message Recorder still contains
untransmitted data. Press F6 to start a new recording. A beep will sound
from the recorder.
Deleting Old Data
Press the F6 button to delete any old data remaining on the recorder and
to begin entering patient data for this test.
Entering Patient Information
The controller displays the patient information screen when you press
the POWER button. Use this screen to enter the followingpatient
information:
ID number (up to 10 digits)
Age
Gender
Last Name [Displayed as NAME(L)]
First Name [Displayed as NAME(F)]
Items can be selected by pressing the buttons.