Product Manual

5-4 SEER
®
Light 2040604-012D
Recording Data
The Preview screen is displayed on the controller, showing the
following elements:
The patient’s heart rate is displayed in the upper right corner
Two or three channels of waveform data are displayed
Adjust patient hookup if waveform height/quality are not
satisfactory or if messages are displayed instead of waveforms.
CAUTION
WAVEFORM DIAGNOSTIC QUALITY — The ECG
waveforms shown on the preview screen are used for
assessing the quality of the connections only.
Do not use these preview waveforms for diagnostic
purposes.
CAUTION
WAVEFORM DISPLAY — Fluorescent lights in the
room may interfere with infrared communication between
the recorder and controller. In this case, normal ECG
waveform will not be displayed on the LCD.
Keep the device away from fluorescent lights. Turn off
fluorescent lights when device is used. Cover the infrared
window with a hand to shield the display from the
fluorescent lights.
4. Select the F4 button to exit the Preview screen and return to the
SEER screen.