Product Manual
5-6 SEER
®
Light 2040604-012D
Recording Data
1. Select the SEER Light Hookup icon on your desktop. The GE
SEER Light Hookup window appears.
2. Connect the USB cable to the connect and the Holter analysis PC.
When properly connected, the LED on the connect turns amber.
NOTE
The following picture shows a USB port on the front of the PC.
Some PCs have the USB port(s) located on the back. The connect
can be connected to either a front or back USB port.
086A
3. Align the infrared terminals on the connect and the recorder.
4. At the GE SEER Light Hookup window, select the Start Preview
button.
Up to three channels of waveform data are displayed.
Adjust patient hookup if waveform height/quality are not
satisfactory or if messages are displayed instead of waveforms.
CAUTION
WAVEFORM DIAGNOSTIC QUALITY — The ECG
waveforms shown on the preview screen are used for
assessing the quality of the connections only.
Do not use these preview waveforms for diagnostic
purposes.
5. Select the Stop Preview button to stop the lead preview.
Connect square
USB connection
here.
When the LED turns
amber, the USB
cable is properly
connected.
Connect rectangular
USB connection to
the computer’s USB
port.