Product Manual
5-8 SEER
®
Light 2040604-012D
Recording Data
Considerations Using the SEER Light Connect
Insert the batteries in the recorder before entering the preview mode.
The ECG waveform cannot be checked without inserting the
batteries.
By entering the preview mode, infrared data transfer will be checked
automatically.
You can only enter preview mode by remote wireless operation before
starting the recorder (selecting the Record button) and during the
first three minutes of the recording phase (after pressing the Record
button). The recorder will beep for the first minute of the recording
phase.
The recorder will enter Power Save mode after setting the batteries.
Confirming ECG Recording Using the SEER Light Connect
1. At the GE SEER Light Hookup window, select the Start Preview
button.
NOTE
Selecting the Start Preview button will cause the button text to
change to Stop Preview.
2. At the Lead Preview screen, review the displayed channels for ECG
waveform confirmation.
NOTE
You may continue to enter patient demographics while the
waveform is scrolling.
3. Select the Stop Preview button after checking the ECG waveforms.
Interruption of Infrared Communication
Distance, direction, objects, and so on. between the recorder and
controller or connect can interrupt the infrared communication, causing
the ECG waveform to become a bold line on the LCD display.