Product Manual

5-10 SEER
®
Light 2040604-012D
Recording Data
CAUTION
PACEMAKER PATIENTS — Pacemaker pulses are only
detected on CH1. Before attaching the recorder, check the
patient’s ECG to predict which lead will produce a big
pacemaker pulse. That is the lead that should be assigned
to CH1 (refer to table below).
If the pacemaker pulse of CH1 is small, there is a
possibility that the pacemaker pulse may not be detected.
Change the lead of CH1 to another lead that has a bigger
pacemaker pulse.
The down pointing arrow mark of the pacemaker pulse
location for the preview mode is only displayed when
using the remote method.
To verify accurate pace detection using the connect:
At the Lead Preview window, the position of the pacemaker spikes are
displayed with vertical bars. Refer to the following illustration.
Holter ECG Leads Similar 12-Leads
CM
5
, CC
5
V
5
CM
1
II, III, aV
1
NASA V
1