Product Manual

Recording Data
2040604-012D SEER
®
Light 5-13
CAUTION
BATTERIES — Do not use any battery other than new
alkaline batteries. Use of other types of batteries,
including manganese, rechargeable, or used alkaline
batteries will not ensure ECG recording for the life of the
recorder (24 hours for the SEER Light recorder, 48 hours
with the SEER Light Extend recorder).
Do not leave the batteries in the recorder for a long time.
This may cause leakage of battery liquid.
If the recorder (with batteries inserted) is not started for
over one hour when the power is ON, the battery Power
Save mode will be activated. In this case, the batteries
must be removed to deactivate the Power Save mode. To
reset the recorder, leave the recorder without the
batteries for 30 seconds.
Make sure the batteries are set in the correct directions.
Then begin the recording process again.
Begin Recording with SEER Light Extend Controller
1. Press the F6 button on the controller to start recording. Refer to
“Entering Patient Data” on page 4-1.
NOTE
Recording can be started without entering patient information.
NOTE
If data remains on the recorder that has not been transmitted to
the controller, the message Recorder still contains untransmitted
data. Press F6 to start a new recording. will be displayed, and a
beep will sound from the recorder. Press the F6 button to delete
the data and start recording.
If the data is needed, transfer the remaining data to the
controller. Refer to Chapter 6, “Transferring Data”.
2. Confirm that the beep sounds for one minute and that the LED
flashes (twice per second) for three minutes.
The recording date and time will be displayed on the LCD of the
controller.