Product Manual

5-14 SEER
®
Light 2040604-012D
Recording Data
3. Put the recorder in the carrying pouch with the button facing
up.
Begin Recording with SEER Light Connect
1. Select the SEER Light Hookup icon located on your desktop.
The GE SEER Light Hookup window appears.
NOTE
If data remains on the recorder that has not been transmitted to
the connect, the message Recorder contains untransmitted data.
Press the “Record” button to start a new recording is displayed.
If the data is needed, transfer the remaining data to the connect.
Refer to Chapter 6, “Transferring Data”.
2. Point the infrared terminal of the recorder to the connect.
3. At the GE SEER Light Hookup window, select Record.
A message appears stating the Holter recording started successfully.
Select OK.
NOTE
If the infrared sensors are not aligned properly, or if the cables
are not connected, a message will appear stating Error occurred
while starting the recorder. Check the connections and infrared
sensor alignment.
4. Confirm that the beep sounds for one minute and that the LED
flashes (twice per second) for three minutes.
NOTE
If you press the Record button after the recording has already
started, a message appears stating Holter recording has already
started.
5. Put the recorder in the carrying pouch with the button facing
up.
Begin Recording with SEER Light Recorder Button
1. Insert the batteries. Refer to “Inserting Batteries in the SEER Light
Recorder” on page 5-12 for more information.