Product Manual

6-6 SEER
®
Light 2040604-012D
Transferring Data
Press the ENT button to save the data and return to the screen
in Step 3.
5. Press the F4 button to return to SEER mode.
Displaying the ECG Waveform
1. Insert the flash card into the controller.
NOTE
If the flash card is not inserted into the controller before starting the
operation, error messages will appear on the screen.
2. Press the F3 button under SEER mode. It will proceed to the
confirmation of flash card contents.
3. Select ECG Display. Press ENT.
4. Select the ECG waveform that you want to display by using the
and buttons. Choices are Event, Start, End and Time. Press the
ENT button.
If you select Time, the starting time of the recording will be
displayed first. By entering the time you want to review in this
screen and pressing the ENT button, the ECG waveform
corresponding to that specified time will be displayed on the
screen.
Scroll forward through the ECG waveform by pressing the or
F1 button. You cannot scroll backward through the waveform.
5. Press the or F4 button until you are returned to the screen in Step
4. Press the or F4 button again to return to SEER mode.
Transferring Data to the Holter Analysis System via Flash Card
Follow these steps to transfer data from the flash card to the Holter
analysis system.
1. Insert the flash card into the flash card adapter.
NOTE
ECG data recorded by the recorder can only be analyzed by the
Holter analysis system.
2. Insert the flash card adapter into the acquisition unit. Proceed with
data transfer as outlined in the operator’s manual for the Holter
analysis system.