Troubleshooting
2040604-012D SEER
®
Light A-3
The recorder does not start up when the
controller sends a start up command
through infrared, and Communication Error
is displayed.
(Recorder and controller)
Infrared transfer does not work due to the
distance or an obstruction between the
recorder and controller.
Adjust the position of the infrared terminal
part of the recorder. Set both devices
within 1 meter and adjust the direction so
that the terminal of infrared between the
recorder and controller are aligned. Press
F6 button again.
It has been more than 1 hour after batteries
are inserted.
The recorder automatically cancels start-
ready mode in order to reduce battery
power loss. This function makes the F6
button not work.
Press the F6 button again 30 seconds after
batteries are replaced.
The SEER Light Hookup window displays
a dialog box stating Holter recording has
already started. when the Record button is
selected.
(Recorder and connect)
The recorder is already in the recording
process.
Select OK. The dialog box closes and the
recording continues.
The SEER Light Hookup window displays
a dialog box stating Error occurred while
starting the recorder. Check the
connections and the infrared sensor
alignment. when the Record button is
selected.
(Recorder and connect)
Infrared transfer does not work due to the
distance or an obstruction between the
recorder and connect.
Select the OK button to close the dialog
box. Set both devices within 1 meter and
adjust the direction so that the terminal of
infrared between the recorder and connect
are aligned. Select the Record button
again.
Patient data not available because cables
are not properly connected.
Select the OK button to close the dialog
box. Connect the cables. Select the
Record button again.
The recorder does not start up when the
controller sends a start up command
through infrared, and Recorder still
contains untransmitted data. Press the F6
button to start a new recording. is
displayed.
(Recorder and controller)
Data remains in the recorder and was not
transferred to the controller.
If the data is not needed, press the F6
button. Otherwise, transfer the old data to
the controller.
Trouble Cause Recommendation