Troubleshooting
The SEER Light Hookup window displays
a dialog box stating Recorder contains
untransmitted data. Press the Record
button to start a new recording. when the
Record button is selected.
(Recorder and connect)
The recorder contains untransmitted data. Select the OK button to close the dialog
box. If the data is not needed, select the
Record button again. If the data is needed,
transfer the data before beginning a new
recording.
The recorder starts using the controller
command through infrared. However, the
controller displays Communication Error.
(Recorder and controller)
The controller does not receive a response
back to start recording from the recorder.
Align the infrared terminals then repeat the
process. If the Communication Error
condition persists, check with service, or
open a service call.
Patient data is not transferring between the
recorder and the controller.
(Recorder and controller)
An Extend device is used with a non-
Extend device.
Always transfer data from the SEER Light
Extend recorder to a SEER Light Extend
controller.
Beep sound is heard. However, LED does
not blink.
(Recorder)
The recorder is malfunctioning internally. Check with service, or open a service call.
LED blinking stops before 24- or 48-hour
period.
(Recorder)
May be using rechargeable or manganese
type batteries.
Use new alkaline batteries.
The batteries were removed while
recording.
Do not touch or remove the batteries while
recording.
Damage to unit caused by physical shock. Check with service, or open a service call.
When the F1 button on the controller is
pressed for preview through infrared,
transfer error occurs.
(Recorder and controller)
Infrared transfer does not work due to the
distance or an obstruction between the
recorder and controller.
Set both devices within 1 meter and adjust
the direction so that the terminal of infrared
between the recorder and controller are
directly facing each other. Press the F1
button again.
The F1 button was pressed 1 hour after the
batteries are inserted.
Reset the batteries 30 seconds after the
batteries are removed. Press the F1 button
within 1 hour.
The recorder beeps the first minute after
the recording has started. For the
remaining 2 minutes of calibration, the
recorder’s LED flashes twice per second.
After the beeping sound, the controller
cannot preview using infrared. Otherwise,
the recorder and controller must be
connected by the cable.
Trouble Cause Recommendation