Product Manual
Troubleshooting
2040604-012D SEER
®
Light A-5
The batteries are not inserted in the
recorder.
Insert the batteries and press the F1
button.
The batteries in the recorder do not have
enough power.
Replace the batteries and press F1 button.
The SEER Light Hookup Lead Preview
window displays no waveforms when the
Start Preview button is selected.
(Recorder and connect)
Infrared transfer does not work due to the
distance or an obstruction between the
recorder and connect.
Set both devices within 1 meter and adjust
the direction so that the terminal of infrared
between the recorder and connect are
aligned. Select the Start Preview button
again.
Patient data not available because cables
are not properly connected.
Connect the cables properly. Select the
Start Preview button again.
The notched waveform is displayed while
using infrared preview.
(Recorder and controller/connect)
Infrared transfer does not work due to the
distance or an obstruction between the
recorder and controller/connect.
Set both devices within 1 meter and adjust
the direction so that the terminal of infrared
between the devices are aligned.
When the flash card is set in the slot of the
controller, CARD ERROR Insert the SEER
card properly displays.
(Controller)
The flash card is inserted upside down.
Place the flash card with mark facing
up on the card into the controller. Press the
power switch to on.
The flash card is not completely inserted
into the slot of the controller
Place the flash card all the way into the slot
of the controller. Press the power switch to
on. If the flash card is correctly positioned,
it should not protrude from the side of the
controller.
When flash card is set in the slot of the
controller, CARD ERROR Press F6 to
format the card is displayed on the
controller screen.
(Controller)
The flash card has not been initialized. Follow the initialization instructions
provided by the flash card manufacturer.
CARD ERROR Press F6 to format the card
is displayed on the controller screen when
the data transfer starts.
(Controller)
A different type of card other than the flash
card is inserted.
Insert the correct flash card.
Card contains unanalyzed data. Pressing
F6 will overwrite this data. is displayed on
the controller screen when the data
transfer to the flash card does not start.
(Controller)
The data is not transferred to the Holter
analysis system, and remains on the flash
card.
If necessary, transfer the data to the Holter
analysis system. Otherwise, press the F6
button to overwrite the old data with the
new recording.
Trouble Cause Recommendation