Product Manual

Maintenance
Verifythatallcordsandconnectorsaresecurelyseated.
Inspectkeysandcontrolsforproperoperation:
Togglekeysshouldnotstickinoneposition.
Knobsshouldrotatefullyinbothdirections.
CleaningandDisinfectingExteriorSurfaces
Cleananddisinfecttheexteriorsurfacesofallequipmentandperipheraldevices
monthly,ormorefrequentlyifneeded.
CleaningandDisinfectingtheSurfaces
Propercleaninganddisinfectingprolongsthelifeoftheproduct.Failuretousethe
propercleaningsolutionsortofollowproperprocedurescanresultinthefollowing:
Damageorcorrosion
Productdiscoloration
Metalpartcorrosion
Unitmalfunction
Voidedwarranty
Usethefollowingproceduretocleantheequipment'sexteriorsurfaces.Besureto
observeallcautionswhencleaningthedevice.
1.Toclean,wipewithalightlymoistenedcloth.
Useamildsoapandwatersolutiontomoistenthecloth.
DoNOTuseanyofthefollowingcleaningproducts,orproductsthatcontainthe
sameactiveingredientsandsolutions,whichareknowntocausetheproblems
previouslylisted:
Sani-Cloth®Wipes
Ascepti®Wipes
HBQuat®
Clorox®Wipes(theydonotcontainbleach)
Over-the-counterdetergents(suchasFantastic®,Tilex®,andsoon)
2.Todisinfect,wipewithasoft,lint-freeclothmoistenedwithanappropriate
disinfectant.
Usethefollowingsolutions,asrecommendedintheAPICGuidelinesforSelection
andUseofDisinfectants(1996):
Sodiumhypochlorite(5.2%householdbleach)minimum1:500dilution
(minimum100ppmfreechlorine)andmaximum1:10dilution.
Anysodiumhypochloritewipeproductthatmeetsthepreviousguidelines
canbeused.
3.Drywithacleanclothorpapertowel.
164
MAC
®
5500/MAC
®
5500HD
2046275-016A