Product Manual
RecordinganECG
Promptlevelinthesystemsetup.Ifamessageandpromptisdisplayed,thelead
qualityindicatorwillreecttheoverall10-secondleadquality.
•IfPreviewbeforeanalysisisenabled,thesystemsetupPromptlevelisdisregarded
andthesystemimmediatelydisplaysthePreviewscreen.Anyleadquality
messageswillbedisplayedinthisscreenalongwiththeoverall10-secondlead
qualityindicator.
Ineithercase,usersmaythendoeitherofthefollowing:
•SelectContinuetocontinue(printtheECG).
•SelectCanceltocancel.
StandardECGs
Usethefollowingmethodforresting,pediatric,vectorloop,and15–leadECGs.
1.Afterpreparingthepatient,selectthecorrectECGsoftkey.
•ToconductarestingECG,pressRestingECG.
•ToconductapediatricECG,pressPediatricECG.
•ToconductavectorloopsECG,pressVectorLoops.
•Toconducta15-leadECG,press15leadECG.
TheappropriateECGtestwindowopens.
NOTE:
Alloptionsmaynotbeavailable,dependingonsystemconguration.
2.Enterthecorrectpatientinformation.
RefertoChapter4,“EnteringPatientInformation”fordetails.
3.AdjusttheECGsettingsasappropriate:
•Toselectadifferentsetofleads,pressLeads.
Alistofavailableleadsopens:
•Allleadsselectsalltheavailableleads.
•V1,II,V5selectstheidentiedleads.
•I,II,aVR,aVL,aVFselectstheidentiedleads.
•V1,V2,V3,V4,V5,V6selectstheidentiedleads.
•Leadcheckconductsacheckofallavailableleadsanddisplaystheir
results.
•Leadplacementconductsaleadcheckanddisplaysadiagramofthe
properplacementforeachlead.
•Toadjustthespeedwithwhichthewaveformsaredisplayed,press25mm/s.
Thisopensalistofavailablespeeds.Optionsinclude5mm/s,12.5mm/s,
25mm/s,and50mm/s.Thisaffectsonlythespeedofthewaveformasit
appearsonthescreenandreports;itdoesnotaffecttheactualspeedwith
whichtheECGisrecorded.Theselectedvalueappearsonthefunctionkey
onthescreen.
•ToadjustthegainoftheECG,press10mm/mV.
2046275-016A
MAC
®
5500/MAC
®
5500HD
73