Product Manual

2036991-001C CARESCAPE V100 Vital Signs Monitor 9-9
Nellcor OXIMAX SpO2: Specifications
Specifications
Specifications
Measurement range
SpO
2
1 to 100%
Pulse rate 20 to 250 bpm
Perfusion range 0.03 to 20%
Accuracy
Saturation
Adult* 70 to 100% ±2 digits
Neonate* 70 to 100% ±3 digits
Low perfusion** 70 to 100% ±2 digits
Pulse Rate
Adult and neonate 40 to 250 bpm ±3 digits
Low perfusion** 40 to 250 bpm ±3 digits
*Adult specifications are shown for OxiMax MAX-A and MAX-N sensors with the N-600. Saturation accuracy will vary by sensor type. This
variation equals plus or minus one standard deviation. Plus or minus one standard deviation encompasses 68% of the population. Accuracy is
based on deep hypoxia studies on healthy adult volunteer subjects. Arterial blood samples were analyzed simultaneously on multiple CO-
oximeters
**Applicability: OxiMax MAX-A, MAX-AL, MAX-P, MAX-I, and MAX-N sensors.
Specifications: Nellcor OXIMAX
sensor accuracy
NOTE: All Nellcor OXIMAX
sensors must be used with the NELL cable; the SCP-10 cable. RS-10 and Oxisensor II
sensors are not compatible with the V100 Vital Signs Monitor.
Sensor Model
SpO
2
Range
70% to 100%
OXIMAX
MAX-A, MAX-AL ± 2 digits
MAX-N
(adult)
± 2 digits
MAX-N
*
(neonate)
± 3 digits
MAX-P ± 2 digits
MAX-I ± 2 digits
MAX-FAST ± 2 digits