9-10 CARESCAPE V100 Vital Signs Monitor 2036991-001C
Nellcor OXIMAX SpO2: Specifications
SC-A (adult) ± 2 digits
SC-PR (neonate) ± 3 digits
SC-NEO ± 3 digits
MAX-R
**
± 3.5 digits
OxiCliq
OxiCliq A ± 2.5 digits
OxiCliq P ± 2.5 digits
OxiCliq N
(adult)
± 2.5 digits
OxiCliq N
*
(neonate)
± 3.5 digits
OxiCliq I ± 2.5 digits
Reusable sensor models
D-YS (infant to adult) ± 3 digits
D-YS (neonate) ± 4 digits
D-YS & D-YSE ± 3.5 digits
D-YS & D-YSPD ± 3.5 digits
DS-100A ± 3 digits
OXI-A/N (adult) ± 3 digits
OXI-A/N (neonate) ± 4 digits
OXI-P/I ± 3 digits
Neonatal sensor accuracy When sensors are used on neonatal subjects as
recommended, the specified accuracy range is increased
by ±1 digit,
as compared to adult usage, to account for
the theoretical effect on oximeter measurements of fetal
hemoglobin in neonatal blood. For example, MAX-N
accuracy on neonates is ±3 digits, rather than ±2 digits.
Sensor light source
Wavelength*** Infrared: 890 nm (nominal)
Red: 660 nm (nominal)
Power dissipation Infrared: 22.5 mW (max)
Red: 30 mW (max)
*
The MAX-N, D-YS, OXI-A/N, and OxiCliq N were tested on patients >40 kg.
**
The accuracy specification has been determined between saturations of 80%-100%.
*** Information about wavelength range can be especially useful to clinicians.
Specifications: Nellcor OXIMAX
sensor accuracy