2036991-001C CARESCAPE V100 Vital Signs Monitor 9-15
Nellcor OXIMAX SpO2: Troubleshooting
PROBLEM: An E20 SENSOR DISCONNECTED error code has been detected, but
the sensor is still connected.
CAUSE:
The sensor is not completely connected. The sensor extension cable or
sensor wiring is faulty.
SOLUTION:
Check the patient.
Check the sensor connection to the sensor extension cable and the sensor
extension cable connection to the monitor. Then, if needed, replace the
sensor or the sensor extension cable.