Product Manual
2036991-001C CARESCAPE V100 Vital Signs Monitor 10-3
Masimo SET SpO2: Description
Description
The SpO
2
parameter in the V100 Monitor is available in three different leading
technologies: Ohmeda TruSignal, Nellcor and Masimo SET. Please refer to the
front of your monitor to see which SpO
2
technology your monitor contains. If
your monitor contains this parameter, the SpO
2
technology logo will be on the
front fascia of the monitor. This section refers to Masimo SET SpO
2
technology.
Functional oxygen saturation (SpO
2
) of arterial blood is noninvasively and
continuously monitored in the V100 Monitor using pulse oximetry technology
from Masimo SET. Functional SpO
2
is the ratio of oxygenated hemoglobin to
hemoglobin that is capable of transporting oxygen. This ratio, expressed as a
percentage, is shown in the SpO
2
window, which is continually updated.
Pulse rate when associated with SpO
2
appears in the Pulse Rate window. A tone
sounds at a rate corresponding to the pulse rate and at a pitch corresponding to
the SpO
2
saturation level. The pitch is highest at 100% oxygen saturation, and it
becomes lower as the saturation level falls. The monitor displays a pulse
amplitude bar. The pulse amplitude bar graph is proportional to the arterial
blood flow.
The parameter automatically switches on when a sensor is connected to the
monitor.
Audible and visible alarms occur when SpO
2
levels are outside the alarm limits.
When a parameter status alarm code occurs, an alarm message appears in the
SpO
2
window.
NOTE: Limit alarms, printing, and trending are not available for the first 10
seconds of SpO
2
monitoring.
Indications and Contraindications
The SpO
2
parameter is indicated for use in continuous, noninvasive monitoring
of functional oxygen saturation and in providing pulse rate data as a component
of the CARESCAPE V100 Vital Signs Monitor. This device is not designed, sold, or
intended for use except as indicated.
®
Masimo SET
®
, LNOP, and LNCS are trademarks of Masimo Corporation. Possession or purchase of this
device does not convey any express or implied license to use the device with replacement parts
which would, alone, or in combination with this device, fall within the scope of one or more of the
patents relating to the device.