10-6 CARESCAPE V100 Vital Signs Monitor 2036991-001C
Masimo SET SpO2: Configuration Settings Associated with SpO2
NOTES
A patient’s vital signs may vary dramatically during the use of
cardiovascular agents such as those that raise or lower blood pressure or
those that increase or decrease pulse rate.
The V100 Vital Signs Monitor that is labeled with Masimo SET Technology is
compatible only with Masimo SET sensors.
Software development, software validation, and Risk and Hazard Analysis
has been performed to a registered quality system.
User or patient-applied parts are latex-free.
Configuration Settings Associated with SpO
2
There are three configuration settings associated with this parameter: averaging
time (
n0d), FastSAT (Sat), and sensitivity mode (Sen).
Averaging Time
(
n
0d)
allows you to choose the number of seconds over which SpO
2
data is averaged. Choose
4
to
16
in steps of 2 (alarms are delayed by this amount).
NOTE: Increased averaging time effects time to alarm for saturation and pulse
rate limits.
FastSAT (Sat) allows you to choose 0 (for Off) or 1 (for On). If FastSAT is configured
to 1 (On), the SpO
2
values are calculated quicker.
Sensitivity mode (Sen) setting allows you to adjust the thresholds for calculating
SpO
2
values under low perfusion conditions. Choose 1 (low perfusion-
Maximized), 2 (low perfusion-Default), or 3 (adaptive probe off).
NOTE: Adaptive probe off provides a mode with enhanced detection of
"probe off " conditions. It is intended to be used if normal mode is
not detecting "probe off" with some sensors and conditions.
Buttons Associated with SpO
2
There are no buttons associated with this parameter.
Windows Associated with SpO
2
There is one window associated with this parameter: SpO
2
. While in offline
mode, nothing is displayed in the SpO
2
window. When in ready mode and the
parameter senses that a sensor is connected, a single dash (-) appears in this
window. When it is in operate mode and the parameter is reporting valid data,
the derived SpO
2
value appears in this window and updates continuously. The
values are displayed in %.
NOTE: If SpO
2
is the source for pulse rate, the Pulse Rate window is associated
with this parameter. Refer to the “Pulse Rate” Section for more
information.