2036991-001C CARESCAPE V100 Vital Signs Monitor 10-7
Masimo SET SpO2: Indicators Associated with SpO2
Indicators Associated with SpO
2
There is one indicator associated with this parameter: the pulse amplitude
indicator bar. The red LED bar flashes to indicate that pulse rate measurements
are being derived from SpO
2
signals and the height of the bar is proportional to
the arterial blood flow.
User Settings
Limit settings
There are two limit settings associated with this parameter: HIGH and LOW. The
range for HIGH is 71 to 100%. The range for LOW is 70 to 99%. The settings
appear in increments of 1%.
Menu Settings
The PULSE VOLUME menu setting is associated with the SpO
2
parameter. This
option lets you adjust the volume of the tone that sounds for each pulse
detected. It can be adjusted from 0 to 10 (10 being the loudest). If you set the
volume to zero, no tone will sound.
Sounds Associated with SpO
2
An audible tone is provided by the monitor for each pulse detected by the
Masimo SET SpO
2
parameter. The pitch of the audible tone is directly related to
the calculated saturation value. As the saturation value increases, the pitch
frequency increases. As the saturation value decreases, the pitch frequency
decreases. This audible tone is silenced while an alarm sounds or the PULSE
VOLUME is set to 0. Refer to “Menu Settings” in this section.
Procedures
1. Select a sensor that is appropriate for the patient and the clinical situation.
WARNING
Do not use a damaged sensor or one with exposed electrical
contacts.
NOTE: Use only Masimo sensors, which are available from Masimo
Corporation and GE Medical Systems Information Technologies-
Accessories and Supplies.