2036991-001C CARESCAPE V100 Vital Signs Monitor 10-9
Masimo SET SpO2: Alarms
Alarms
SpO
2
Hold-off Period
The hold-off period is defined as the 10 seconds that follows when the SpO
2
parameter switches from ready mode to operate mode. This hold-off period
helps prevent any nuisance alarms.
Parameter status alarms are not displayed and the derived SpO
2
value is not
checked against user-set limits while the SpO
2
hold-off period is active.
Alarm Timer
When an SpO
2
sensor is on for less than 2 minutes this is considered "spot
mode." The SpO
2
--- SENSOR OFF FINGER and E25 SpO
2
LOST PULSE alarms are
generated as low-priority alarms if they occur within the spot mode time. While
in spot mode, the length of spot mode operation is further increased by taking a
manual NIBP determination.
Under all other conditions, these alarms are generated as high-priority alarms. If
the low-priority alarms are not acknowledged within 1 minute, they escalate to
high-priority alarms.