Product Manual

2036991-001C CARESCAPE V100 Vital Signs Monitor 10-15
Masimo SET SpO2: Troubleshooting
Troubleshooting
This section discusses potential difficulties and suggestions for resolving them. If
the difficulty persists, contact a qualified service person or your local
representative.
The service manual, which is for use by qualified service personnel, provides
additional troubleshooting information.
PROBLEM: The pulse amplitude bar indicates a pulse, but no oxygen saturation
or pulse rate values appear on the screen.
CAUSE:
Excessive patient motion may be making it impossible for the SpO
2
function to find a pulse pattern.
The sensor may be damaged.
The patient’s perfusion may be too low to allow the SpO
2
function to
measure saturation and pulse rate.
SOLUTION:
Check the patient.
If possible, keep the patient still; check whether the SpO
2
sensor is
applied securely and properly, and replace it if necessary; move the
sensor to a new site; or use a disposable adhesive sensor that may
tolerate more motion.
Replace the sensor.
PROBLEM: The SpO
2
value or the pulse rate changes rapidly; the pulse amplitude
bar is erratic.
CAUSE:
Excessive patient motion may be making it impossible for the SpO
2
function to find a pulse pattern.
An electrosurgical unit (ESU) may be interfering with performance.
SOLUTION:
Check the patient.
If possible, keep the patient still; check whether the sensor is applied
securely and properly, and replace it if necessary; move the sensor to a
new site; use a sensor that tolerates more motion.
If an ESU is interfering:
Move the SpO
2
cable as far from the ESU as possible.
Plug the monitor and the ESU into different AC circuits.
Move the ESU ground pad as close to the surgical site as possible.
The sensor may be damp or may need to be replaced with a new
sensor.