Product Manual
2036691-001C CARESCAPE V100 Vital Signs Monitor 11-11
Alaris Turbo Temp: Specifications
Specifications
Factory Default Settings
Unit of measure: °F
Alaris
Patents
U.S. D300,728, D300,909. Other pending patents.
Specifications
Scale °Fahrenheit (F)°
°Celsius (C)
Range
Predictive mode Max: 41.1°C; 106.0°F
Min: 35.6°C; 96.0°F
Monitor mode Max: 41.1°C; 106.0°F
Min: 26.7°C; 80.0°F
Monitor mode accuracy ±0.1°C
±0.2°F (when tested in a calibrated liquid bath; meets ASTM
E1112, Table 1, in range specified)
Determination time approx. 10 seconds, typical
NOTE: Use only IVAC probes and P850A probe covers. The size, shape, and thermal characteristics of the probe covers can affect the
performance of the instrument. Inaccurate readings or retention problems may occur unless IVAC probes and probe covers are used.
Refer to Appendix D for reorder codes.