Product Manual

13-4 CARESCAPE V100 Vital Signs Monitor 2036991-001C
Battery: Buttons Associated With the Battery
Buttons Associated With the Battery
There are no buttons associated with the battery.
Windows Associated With the Battery
There are no windows associated with the battery.
Indicators Associated With the Battery
When the monitor is on, the DC charger is not attached, and the battery is
sufficiently charged, the BATTERY OK indicator is backlit green. Unless specified
otherwise, at all other times this indicator is turned off.
When the BATTERY LOW alarm is active as a low-priority alarm, the BATTERY
LOW indicator is backlit amber and does not flash. When the BATTERY LOW
alarm is active as a high-priority alarm, the BATTERY LOW indicator flashes
amber according to the alarm duty cycle for high-priority alarms.
The CHARGING indicator is backlit green whenever DC charger is attached to
the monitor. Unless specified otherwise, at all other times this indicator is turned
off.
NOTE: Refer to “Alarms Section below for more information.
First Use
To condition a new sealed lead acid battery and optimize its performance, plug
in the monitor; the internal battery pack then charges automatically. Before the
V100 Monitor is used for the first time, the battery should be charged in the
monitor for at least 8 hours. With external DC power connected, the green
CHARGING indicator will light to indicate that the battery is charging. Prior to
first use and after 8 hours of charging, be sure the BATTERY OK indicator is
backlit green when the charger is not connected and the monitor is on.
Battery Charging
Prior to each use, inspect the power supply cord to ensure proper connection
and condition.
With external DC power connected, the green CHARGING indicator will light to
indicate that the battery is charging. This indicator remains active whether the
unit is on or off. An audible "two beep" sounds whenever the DC charger is
connected/disconnected.