13-6 CARESCAPE V100 Vital Signs Monitor 2036991-001C
Battery: Alarms
Alarms
Battery Low
When about 45 minutes of battery charge remains:
The low-priority BATTERY LOW alarm is issued.
The BATTERY LOW indicator illuminates.
This alarm can be silenced by pressing the Silence button.
The BATTERY LOW alarm will re-alarm every 10 minutes after it’s been
silenced.
If the alarm is not silenced, the alarm is re-issued every 8 seconds.
The monitor continues to operate normally.
When about 5 minutes of battery charge remains:
The low-priority BATTERY LOW alarm escalates to a high-priority BATTERY LOW
alarm.
The BATTERY LOW indicator flashes.
Any NIBP determination in progress at the time of the alarm escalation
is allowed to finish.
Any Stat mode cycle that was initiated before the alarm escalation is
allowed to finish.
The user is not able to initiate:
any new NIBP determinations of any type
any printouts
NOTE: At this time, it is highly recommended to plug the monitor into external
DC power.
When 5 minutes of battery charge expires:
After 5 minutes of high-priority BATTERY LOW alarm, the monitor enters a
battery low shutdown.
No error code is displayed.
The BATTERY LOW indicator flashes.
The monitor alarms for 2.5 minutes then shuts down completely.
CAUTION
You must plug the monitor into DC power before resuming
monitoring.