C-4 CARESCAPE V100 Vital Signs Monitor 2036991-001C
Maintenance: Maintenance, Calibration, and Cleaning
CAUTIONS
Never immerse monitor or accessories in any liquid.
Do not attach the monitor or accessories to a patient until it is
thoroughly dry.
WARNING
Failure to follow these cleaning recommendations may melt,
distort, or dull the finish of displays and cases; blur lettering on
labels; embrittle cases and lead to cracks and breakage; or
cause equipment failures. Use of non-approved cleaning
agents is not considered normal wear and repair or
replacement of parts is not covered under warranty.
Monitor Exterior
Disconnect the monitor from AC power before cleaning or disinfecting its
surface. The exterior surfaces of CARESCAPE Monitors may be cleaned with a
dampened, lint-free cloth. Wipe off all cleaning solutions with a clean, dry cloth
and let air dry for at least 15 minutes. Use one of the following approved
solutions:
Mild soap (diluted)
Commercial diluted bleach solution or bleach wipe
Commercial diluted ammonia solution
10% solution of household bleach (5.25% sodium hypochlorite) in distilled
water
Never use the following cleaning agents: –
Abrasive cleaners or solvents of any kind
Acetone
Ketone
Betadine
Alcohol- or petroleum-based cleaning agents
Any type of solution that contains ammonium chloride, conductive
solutions, wax or wax compounds
Sodium salts
NOTE: Never autoclave or steam clean the monitor, cuffs, or accessories.
Monitor Display
To clean the display screen, use a soft, clean cloth dampened with a glass
cleaner. Never spray the glass cleaner directly onto the display, and never use
alcohol- or petroleum-based products.