Product Manual
2036991-001C CARESCAPE V100 Vital Signs Monitor C-7
Maintenance: Battery and Storage Care
CAUTIONS
To ensure that the battery will be ready for portable operation,
keep the monitor connected to a mains supply whenever
possible.
Repeated failure to fully charge the battery will result in a
significant reduction in battery life.
The expected lifetime of the battery largely depends on the
way in which the monitor is used. If the battery is allowed to
completely discharge before being fully recharged, the battery
should survive around 450 charge/discharge cycles.
Replacing the Battery
NOTE: Record the configuration settings on your monitor before replacing the
battery. When the battery is replaced, all user settings are lost and
return to default values.
1. Unplug the monitor from the DC power source.
2. Looking at the bottom of the V100 Monitor, remove the battery
compartment cover by removing the four screws that secure the cover and
help card tray.
3. Remove the help card tray and battery door cover.
4. Remove the old battery and disconnect the wires. Attach the battery wires
to the new battery, ensuring the red terminal (+) is connected to the red wire
and the black terminal (-) is connected to the black wire.