2036991-001C CARESCAPE V100 Vital Signs Monitor 2-7
Getting Started: Configuration Mode Settings
Configuration Mode Settings
Monitor settings such as HIGH/LOW alarm settings changed in the clinical mode
will not be retained after the monitor is powered off. To retain alarm and
parameter settings, the changes must be done in the configuration mode. Date/
Time settings are also entered in the configuration mode.
Entering Configuration Mode
1. With the monitor off, press and hold the Menu button at the same time as
pressing the On/Off button until the display test completes.
NOTE: CFG displays in the Systolic window.
As the monitor turns on in the configuration mode, a brief display appears
showing the software revision and the NIBP technology of the monitor. These
displays appear only during the first part of the power up sequence and are not
selectable and cannot be changed.
The Menu selections appear in the following order. Refer to each manual section
for settings options.
Menu selections for SpO
2
differ depending upon the technology. Refer to the
“Ohmeda TruSignal SpO
2
,”
“Masimo SpO
2
,” and “Nellcor SpO
2
,” Sections for
options.
Display Window
Major software revision Systolic
Minor software revision Diastolic
Type of NIBP technology min
Setting Setting LED Window LED Display
Inflate pressure (adult/ped) Systolic XXX (numeric)
Inflate pressure (neonate) Systolic XXX (numeric)
Line frequency mode
(Ohmeda TruSignal only)
SpO
2
SpO
2
mode
(Nellcor & Masimo only)
SpO
2
SpO
2
sat
(Nellcor & Masimo only)
SpO
2
SpO
2
sensitivity
(Masimo only)
SpO
2