Product Manual
2036991-001C CARESCAPE V100 Vital Signs Monitor 2-9
Getting Started: SpO2 Configuration Settings
SpO
2
Configuration Settings
Procedure for Units With Ohmeda TruSignal Technology
(Refer to the “Ohmeda TruSignal SpO
2
” Section for options)
1. With the monitor off, press and hold the Menu button at the same time as
pressing the On/Off button until the display test completes.
2. Press the Menu button until LF appears in the Pulse Rate window.
3. Use the +/- buttons to select the option.
4. To exit the configuration mode, turn the unit off. To continue with additional
configuration settings, press the Menu button.
Procedure for Units With Nellcor Technology
(Refer to the “Nellcor SpO
2
” Section for options)
1. With the monitor off, press and hold the Menu button at the same time as
pressing the On/Off button until the display test completes.
2. Press the Menu button until n0d (response mode) appears in the Pulse Rate
window.
3. Use the +/- buttons to select the option.
4. Press the Menu button once. SAt (SatSeconds) appears in the Pulse Rate
window.
5. Use the +/- buttons to select the option.
6. To exit the configuration mode, turn the unit off. To continue with additional
configuration settings, press the Menu button.
Procedure for Units With Masimo Technology
(Refer to the “Masimo SpO
2
” Section for options)
1. With the monitor off, press and hold the Menu button at the same time as
pressing the On/Off button until the display test completes.
2. Press the Menu button until n0d (averaging time) appears in the Pulse Rate
window.
3. Use the +/- buttons to select the option.
4. Press the
Menu
button once.
SAt
(FastSAT) appears in the
Pulse Rate
window.
5. Use the +/- buttons to select the option.
6. Press the Menu button once. SEn (sensitivity mode) appears in the Pulse
Rate window.
7. Use the +/- buttons to select the option.
8. To exit the configuration mode, turn the unit off. To continue with additional
configuration settings, press the Menu button.