Product Manual
2036991-001C CARESCAPE V100 Vital Signs Monitor 3-5
Product Overview: Rear Panel
27. AUTO CYCLE indicator: lights green to indicate auto mode is the chosen
NIBP mode; flashes to indicate you are making a change to the auto mode.
28. Min window: displays the NIBP mode if manual or Stat as well as the cycle
time when taking auto NIBP determinations.
29. HISTORY indicator: flashes to indicate you are in history mode.
30. BATTERY OK indicator: lights green to indicate the monitor is operating on
battery power and that the battery is sufficiently charged.
31. BATTERY LOW indicator: lights amber to indicate low charge for the battery
(45 min or less when solid; 5 min or less when flashing).
32. CHARGING indicator: lights green to indicate presence of external power
source and battery charging.
33. Temperature window: lights 4-digit red LED to indicate measured
temperature.
Rear Panel
34. Data interface connector: host communications port (15 pin D-type RS-232
serial port) for use only with equipment conforming to IEC 60601-1,
configured to comply with IEC 60601-1-1.
35. Printer door.
34
35