2036991-001C CARESCAPE V100 Vital Signs Monitor 3-9
Product Overview: User Modes
Menu Mode
The menu mode allows you to access and change the three settings associated
with the following indicators: INFLATE PRESSURE (ADULT and NEONATE),
ALARM VOLUME, and PULSE VOLUME.
To enter this mode, press the Menu button. Each press of the Menu button steps
you through each of the these settings.
After 7 seconds of not pressing the Menu button, the menu mode is
automatically exited. Otherwise, you can exit the menu mode by cycling through
all menu options. Upon exiting menu mode, the main monitoring screen is
displayed. Alarm and pulse volume settings are retained after power-off.
INFLATE PRESSURE (ADULT, NEONATE) is reset to its configured default after
power-off.
Inflate Pressure
Procedure
NOTE: This setting is available for two patient types: adult and neonate. The
adult setting is applicable to both adult and pediatric determinations.
1. Press the Menu button. The INFLATE PRESSURE indicator flashes, and—at
the same time—the ADULT indicator and the value in the Systolic window
light showing you that the INFLATE PRESSURE for ADULT setting is ready to
be changed.
2. To change the associated value, simply use the +/- button to increment or
decrement, respectively.
3. Press the Menu button again. The INFLATE PRESSURE indicator flashes,
and—at the same time—the NEONATE indicator and the value in the
Systolic window light showing you that the INFLATE PRESSURE for
NEONATE setting is ready to be changed.
4. To change the associated value, simply use the +/- button to increment or
decrement, respectively.
Alarm Volume
Procedure
1. Press the Menu button. The ALARM VOLUME indicator flashes.
2. To change the associated value, simply use the +/- button to increment or
decrement, respectively.
Pulse Volume
Procedure
1. Press the Menu button. The PULSE VOLUME indicator flashes.
2. To change the associated value, simply use the +/- button to increment or
decrement, respectively.