Product Manual
3-10 CARESCAPE V100 Vital Signs Monitor 2036991-001C
Product Overview: User Modes
Cycle Mode
The cycle mode allows you to start auto cycle and Stat modes.
1. Press the Cycle button. The AUTO CYCLE indicator flashes.
2. To change the time increment while the AUTO CYCLE indicator flashes,
simply use the +/- button to increment or decrement, respectively. When
you reach the beginning or ending of the list, the negative key-click sounds.
OR
3. While the AUTO CYCLE indicator flashes, you can also press the Cycle
button until you reach the desired time increment.
Refer to the “NIBP” Section for more information.
Limit Adjustment Mode
The limit adjustment mode allows you to change alarm limit settings that are
used while monitoring a patient. To enter this mode, press the Alarms button. All
alarm limit settings return to their default settings after power-off. To change the
associated limit, simply use the +/- button to increment or decrement,
respectively. The range and increment/decrement steps for each derived vital
sign that has adjustable limits are described in each parameter section. The step
size specified (which cannot be adjusted) tells how much the limit value will
change per increment/decrement key press and also dictates how close
together a pair of limits can be.
Limit-adjustable vital signs are displayed in the following order:
ADULT:
Systolic HIGH, LOW
Diastolic HIGH, LOW
NEONATE:
Systolic HIGH, LOW
Diastolic HIGH, LOW
Pulse Rate:
HIGH, LOW
SpO
2
:
HIGH, LOW
NOTES
The temperature and MAP (mean arterial pressure) vital signs are not
checked against alarm limits.
Only NIBP limits (Systolic and Diastolic) are adjustable based on the patient
type.