Product Manual

2036991-001C CARESCAPE V100 Vital Signs Monitor 3-13
Product Overview: Specifications
Specifications
Specifications
Mechanical
Dimensions
Height 7.7 in (19.5 cm)
Width 8.6 in (21.9 cm) without temperature
10.0 in (25.4 cm) with temperature
Depth 5.3 in (13.5 cm)
Weight (Including battery) 5.4 lb (2.4 kg)
Mountings Self-supporting on rubber feet or pole mounted
Portability Carried by recessed handle
Power requirements
Power converter universal P/N: 2018859-001
Protection against electrical shock Class II
AC input 100 to 250VAC, 12VA
DC output voltage 12VDC at 1A
The AC mains power adapter contains a nonresettable and
nonreplaceable fuse.
Monitor
Protection against electrical shock Internally powered or Class II when powered from specified external
power supply.
DC input voltage 12 VDC, supplied from a source conforming to IEC 60601-1.
Fuses The monitor contains three fuses. The fuses are mounted within the
monitor. The fuses protect the low voltage DC input, the battery, and the
remote alarm output. The +5 V output on the host port connector is
regulated by internal supply.
Battery Refer to “Battery” Section
Environmental
Operating temperature + 5°C to + 40°C
(+ 41°F to + 104°F)
Operating atmospheric pressure 700 hPa to 1060 hPa