4-4 CARESCAPE V100 Vital Signs Monitor 2036991-001C
Printer: Printouts
The availability of the printer is determined at the time the printout is started.
When the printer is unavailable:
The Print button makes a negative key sound when you press it.
Printouts of any type are not available and a high-priority E13 BATTERY
LOW alarm sounds when the BATTERY LOW indicator flashes.
The printer is unavailable if it is out of paper, too hot, or if the monitor is in
any of the following
modes: cycle, alarm limit adjustment, menu, config, or
service.
Printouts
Current (Real Time)
For this printout, the following information may be printed:
SpO
2
info line:
The time the Print button was pressed.
The displayed SpO
2
and pulse rate values are printed under the SpO
2
and pulse rate columns along with the time stamp.
If values are not displayed, “---” is printed.
PIr info line:
The time the Print button was pressed—ONLY if the monitor is
configured for TruSignal SpO
2.
The perfusion index measurement is printed when it is valid. Dashes are
printed when it is invalid (the sensor is not applied to the patient).
NIBP info line:
The displayed NIBP values and the time that these values were
completed.
The displayed pulse rate values if they were completed at the same
time as the displayed NIBP values.
Temperature info line:
The values of a previous temperature measurement if they are still
displayed in the Temperature window.
The time that the measurement completed.
The above lines are printed in the order of most recent to oldest with the
exception of the PIr info line, which always follows the SpO
2
info line. If the
date changes between entries, a single line containing the date is printed.