2036991-001C CARESCAPE V100 Vital Signs Monitor 5-7
Alarms: Alarms
Alarms
Alarm
message code
(if any)
Alarm detected Cause
Alarm
category
Acknowledgeable
by pressing
Silence?*
NIBP
NIBP SYSTOLIC HIGH Value is greater than the HIGH alarm
limit
limit yes
NIBP SYSTOLIC LOW Value is less than the LOW alarm limit limit yes
NIBP DIASTOLIC HIGH Value is greater than the HIGH alarm
limit
limit yes
NIBP DIASTOLIC LOW Value is less than the LOW alarm limit limit yes
E89 NIBP NO DETERMINATION NIBP failed. Reapply cuff status yes
E85 NIBP LEVEL TIMEOUT Motion or irregular pulse. Reduce motion status yes
E84 NIBP TOTAL TIMEOUT Motion or irregular pulse. Reduce motion status yes
E83 NIBP PUMP TIMEOUT Pressure leak. Check or replace hose or
cuff
status yes
E82 NIBP EXCESS AIR IN CUFF Excess amount of air in the cuff. Allow
cuff to deflate
status yes
E80 NIBP OVERPRESSURE Excess cuff pressure. Check for hose
blockage
status yes
SpO
2
SpO
2
HIGH Value is greater than the HIGH alarm
limit
limit no
SpO
2
LOW Value is less than the LOW alarm limit limit no
E20 SpO
2
SENSOR
DISCONNECTED
Sensor disconnected status yes
E21 SpO
2
REPLACE SENSOR Sensor broken or wrong type. Replace status yes
--- SpO
2
SENSOR OFF FINGER Sensor off finger status yes
E25 SpO
2
LOST PULSE Lost pulse status yes
Temperature
E61 TEMP PROBE BROKEN Probe broken. Replace status no
E63 TEMP DISCONNECTED Disconnected or wrong probe status yes
E66 TEMP PROBE TOO HOT Probe too hot status yes