NIBP: Windows Associated with NIBP
Pressing this button key while the BATTERY LOW alarm is active as a high-
priority alarm results in the generation of the E13 BATTERY LOW alarm and the
sounding of the negative key tone.
Windows Associated with NIBP
The windows associated with NIBP are Systolic, Diastolic, MAP/Cuff, Pulse
Rate, and min. The Systolic, Diastolic, MAP/Cuff, and Pulse Rate (if SpO
2
is not
active) windows are automatically cleared when a new NIBP determination is
started. In manual mode, the displayed information is also cleared when it
becomes older than 30 minutes.
The Systolic and Diastolic windows display values after a determination has
completed successfully. While in Stat mode, the Systolic window flashes the
early systolic value if it is available.
The MAP/Cuff window displays the derived mean arterial pressure (MAP)
following the completion of a successful determination. During any type of NIBP
determination, the pressure inside the cuff appears in this window.
The Pulse Rate window displays the NIBP-derived pulse rate when SpO
2
is
inactive.
The min window displays the NIBP mode of operation and the age of the
previous NIBP determination. When both types of information are present, they
flash alternately in this window. When in manual mode, two dashes (- -) are
displayed. When in auto cycle mode, the chosen Cycle time is displayed (e.g., 15).
When in Stat mode, Stat is displayed. When displayed, the age of the previous
NIBP determination is preceded by a minus sign (e.g., - 5 for a determination that
was taken 5 minutes ago).
Indicators Associated with NIBP
The indicators associated with NIBP are Systolic HIGH and LOW, Diastolic HIGH
and LOW, AUTO CYCLE, INFLATE PRESSURE, ADULT, NEONATE, and HISTORY.
The AUTO CYCLE indicator appears solid green when auto mode is on. It flashes
green when changes are being made to the current NIBP mode (e.g., cycle mode
is active). The ADULT indicator appears solid green while systolic and diastolic
limits and INFLATE PRESSURE for adult/pediatric are being adjusted. The
NEONATE indicator appears solid green while limits for systolic and diastolic or
INFLATE PRESSURE for neonate are being adjusted. The HISTORY indicator
appears solid green when the age of the previous NIBP determination is
displayed in the min window.
NOTE: The ADULT indicator encompasses both adult and pediatric