Product Manual
2036991-001C CARESCAPE V100 Vital Signs Monitor 7-9
NIBP: User Settings
cycle mode is deactivated. Press the Inflate/Stop button to stop the
determination in progress without canceling the auto cycle mode. Choose the
two dashes (- -) to cancel auto cycle mode.
If an auto cycle determination results in a limit alarm, a repeat determination is
taken to verify the alarm. Only the first determination in a series of limit alarms
will be followed by a repeat determination.
Whenever an Auto Cycle determination results in an E89 NIBP NO
DETERMINATION alarm, up to nine more repeat determinations are attempted
until valid values are achieved. If at any time during this repeat cycle, the E89
NIBP NO DETERMINATION alarm is silenced by pressing the Silence button or the
Inflate/Stop button, additional determinations are not attempted. If the repeat
cycle completes all nine repeat determinations without reaching a valid value,
the monitor returns to normal auto cycle mode. However, an auto cycle mode
determination must complete successfully before a repeat cycle will follow a
future auto cycle mode determination that results in an E89 NIBP NO
DETERMINATION alarm.
Stat NIBP Determinations
Stat mode allows you to take as many determinations as possible within a 5-
minute time period. The monitor will begin another determination once the
pressure is below 5 mmHg for 8 seconds (neonates) or 15 mmHg for 4 seconds
(adult/pediatric), unless the 5-minute period has ended or Stat mode has been
canceled.
NOTE: NIBP and NIBP-derived pulse rate alarm limits are disabled while in Stat
mode.
Stat NIBP determinations are started by selecting the Cycle button. Once the
Cycle button is pressed, choose Stat. The monitor automatically begins a 5-
minute period of Stat determinations.
NOTE: If the monitor was previously in auto mode, the first Stat NIBP
determination begins after 2 seconds.
After the first Stat determination, subsequent determinations display an early
systolic value that displays in the
Systolic
window. If Stat mode is started when a
determination is already in progress, that determination becomes the first in the
series of Stat determinations. At the end of Stat mode, the NIBP mode prior to
entering Stat mode is resumed. To cancel Stat mode, press the
Inflate/Stop
button.
User Settings
Mode Settings
There is one mode setting associated with this parameter: cycle. The cycle mode
is started by pressing the Cycle button. While the cycle mode is active, cycle
selections are displayed in the min window. Cycle selections appear: Stat, 1, 2, 3,
4, 5, 10, 15, 20, 30, 60, 90, 120, - -. Refer to “Buttons Associated with NIBP” in this
section.