Product Manual
2036991-001C CARESCAPE V100 Vital Signs Monitor 7-15
NIBP: Specifications
Specifications
Specifications
Cuff pressure range
(Normal operating range)
0 to 290 mmHg (adult/ped
0 to 145 mmHg (neonate)
Blood pressure accuracy (Classic and Auscultatory) Meets ANSI/AAMI standard SP-10:1992 (mean error ≤ 5
mmHg, standard deviation ≤ 8 mmHg)
Blood pressure accuracy (SuperSTAT) Meets ANSI/AAMI standard SP-10:2002 (mean error ≤ 5
mmHg, standard deviation ≤ 8 mmHg)
Maximum determination 120 s (adult/ped)
85 s (neonate)
Overpressure cutoff 300 to 330 mmHg (adult/ped)
150 to 165 mmHg (neonate)
BP range (Classic and Auscultatory)
Systolic 30 to 245 mmHg (adult/ped)
40 to 140 mmHg (neonate)
MAP 15 to 215 mmHg (adult/ped)
30 to 115 mmHg (neonate)
Diastolic 10 to 195 mmHg (adult/ped)
20 to 100 mmHg (neonate)
BP range (SuperSTAT)
Systolic 30 to 290 mmHg (adult/ped)
30 to 140 mmHg (neonate)
MAP 20 to 260 mmHg (adult/ped)
20 to 125 mmHg (neonate)
Diastolic 10 to 220 mmHg (adult/ped)
10 to 110 mmHg (neonate)
Pulse rate range (Classic and Auscultatory)
30 to 200 beats/min (adult/ped)
30 to 220 beats/min (neonate)
Pulse rate range (SuperSTAT)
30 to 240 beats/min (adult/ped)
30 to 240 beats/min (neonate)
Pulse rate accuracy ± 3.5% or 3 bpm
NOTE: All CARESCAPE V100 Monitor regulatory and accuracy studies have been performed using GE CRITIKON BP cuffs Use only GE CRITIKON BP
cuffs. The size, shape, and bladder characteristics can affect the performance of the instrument. Inaccurate readings may occur unless
GE CRITIKON BP cuffs are used.