8-10 CARESCAPE V100 Vital Signs Monitor 2036991-001C
Ohmeda TruSignal SpO2: Specifications
Factory Default Settings
GE Medical Systems Information Technologies Patents
6,397,092; 6,748,253; 6,505,133; 7,062,307; 5,766,127; 5,503,148; 5,934,277;
6,385,471; 6,714,803; 6,987,994; 6,408,198; 6,434,408; 6,839,582; 6,505,060;
6,510,329; 6,650,918; 7,139,599; 6,707,257; 6,720,734; 6,825,619 pending.
Specifications: Sensor Accuracy
Sensor model
SpO
2
range
70% to 100%
OxyTip+
OXY-F-UN ±2 digits without motion
OXY-W-UN ±2 digits without motion
OXY-E-UN ±2 digits without motion
OXY-SE ±2 digits without motion
OXY-AP ±2 digits without motion
OXY-AF ±2 digits without motion
OXY-F2-GE ±2 digits without motion
OXY-F4-GE ±2 digits without motion
OXY-E2-GE ±2 digits without motion
OXY-E4-GE ±2 digits without motion
Sensor light source
Wavelength* Infrared: 930 to 950 nm (nominal)
Red 650 to 670 nm (nominal)
Average power < 1 mW
* Information about wavelength range can be especially useful to clinicians.
SpO
2
(%) HIGH alarm limit 100
SpO
2
(%) LOW alarm limit 90
Line frequency mode 60 (for 60 Hz)