Ohmeda TruSignal SpO2: Troubleshooting
PROBLEM: An E20 SENSOR DISCONNECTED error code has been detected, but
the sensor is still connected.
CAUSE:
The sensor is not completely connected. The interconnect cable or sensor
wiring is faulty.
Ensure the appropriate sensor and cable are being used.
SOLUTION:
Check the patient.
Check the sensor connection to the interconnect cable and the interconnect
cable connection to the monitor. Then, if needed, replace the sensor or the
interconnect cable.
Use only compatible sensors and cables.